EasyManuals Logo
Home>Keithley>Multimeter>3700 Series

Keithley 3700 Series User Manual

Keithley 3700 Series
936 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #784 background imageLoading...
Page #784 background image
Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-405
Example 1
timer.reset()
-- (intervening code)
time = timer.measure.t()
print(time)
Resets the timer and measures the time since the
reset.
Output:
1.469077e+01
The output will vary. The above output indicates that
timer.measure.t() was executed 14.69077
seconds after timer.reset().
Example 2
beeper.beep(0.5, 2400)
print("reset timer")
timer.reset()
delay(0.5)
dt = timer.measure.t()
print("timer after delay:", dt)
beeper.beep(0.5, 2400)
Sets the beeper, resets the timer, sets a delay, then
verifies the time of the delay before the next beeper.
Output:
reset timer
timer after delay: 5.00e-01
Also see
timer.reset() (on page 11-405)
timer.reset()
This function resets the timer to zero (0) seconds.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
timer.reset()
Example
timer.reset()
-- (intervening code)
time = timer.measure.t()
print(time)
Resets the timer and then measures the time since
the reset.
Output:
1.469077e+01
The above output indicates that
timer.measure.t() was executed 14.69077
seconds after timer.reset().
Also see
timer.measure.t() (on page 11-404)

Table of Contents

Other manuals for Keithley 3700 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3700 Series and is the answer not in the manual?

Keithley 3700 Series Specifications

General IconGeneral
BrandKeithley
Model3700 Series
CategoryMultimeter
LanguageEnglish

Related product manuals