Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-168 3700AS-901-01 Rev. D/June 2018
dmm.func = "dcvolts"
dmm.reset("active")
dmm.nplc = 0.5
dmm.range = 10
dmm.configure.set("TestDcv")
dmm.configure.recall("dcvolts")
print(dmm.func, dmm.autorange, dmm.range, dmm.nplc)
dmm.configure.recall("TestDcv")
print(dmm.func, dmm.autorange, dmm.range, dmm.nplc)
Set the DMM to the DC volts function.
Reset DC volts back to factory defaults.
Set the NPLC for DC volts to 0.5.
Select the 10 V range for DC volts and disable autorange.
Save a user DMM configuration for DC volts called "TestDcv".
Recall and configure the DMM for factory DC volts.
Output the settings for factory DC volts.
Recall the user DMM configuration called "TestDcv".
Output the settings for the TestDcv configuration.
Output:
dcvolts 1.000000000e+00 1.000000000e+01 1.000000000e+00
dcvolts 0.000000000e+00 1.000000000e+01 5.000000000e-01
Also see
createconfigscript() (on page 11-110)
dmm.configure.catalog() (on page 11-161)
dmm.configure.delete() (on page 11-162)
dmm.configure.query() (on page 11-163)
dmm.configure.recall() (on page 11-165)
dmm.func (on page 11-179)
dmm.setconfig() (on page 11-227)
setup.save() (on page 11-356)
dmm.connect
This attribute indicates how the DMM relays should be connected to the analog backplane.
Reset
DMM reset
Recall setup
Create configuration script
Save setup
Usage
value = dmm.connect
dmm.connect = value
The DMM relay connection setting. See Details for valid values