Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-429
Details
Releases a trigger that was asserted with an indefinite pulse width. It also releases a trigger that was
latched in response to receiving a synchronous mode trigger.
Example
tsplink.trigger[3].release()
Also see
tsplink.trigger[N].assert() (on page 11-424)
tsplink.trigger[N].clear() (on page 11-425)
tsplink.trigger[N].mode (on page 11-426)
tsplink.trigger[N].overrun (on page 11-427)
tsplink.trigger[N].pulsewidth (on page 11-428)
tsplink.trigger[N].stimulus (on page 11-430)
tsplink.trigger[N].wait() (on page 11-431)
tsplink.trigger[N].reset()
This function resets some of the TSP-Link trigger attributes to their factory defaults.
Usage
tsplink.trigger[N].reset()
The trigger line (1 to 3)
Details
The tsplink.trigger[N].reset() function resets the following attributes to their factory
defaults:
â–ª tsplink.trigger[N].mode
â–ª tsplink.trigger[N].stimulus
â–ª tsplink.trigger[N].pulsewidth
This also clears tsplink.trigger[N].overrun.
Example
tsplink.trigger[3].reset()
Resets TSP-Link trigger line 3
attributes back to factory default
values.
Also see
tsplink.trigger[N].mode (on page 11-426)
tsplink.trigger[N].overrun (on page 11-427)
tsplink.trigger[N].pulsewidth (on page 11-428)
tsplink.trigger[N].stimulus (on page 11-430)