Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-328 3700AS-901-01 Rev. D/June 2018
scan.trigger.arm.set()
This function sets the arm event detector to the detected state.
Usage
scan.trigger.arm.set()
Details
This function sets the arm event detector of the trigger model to the detected state.
Example
Sets the arm event detector to the detected
state.
Also see
scan.trigger.arm.clear() (on page 11-327)
scan.trigger.arm.stimulus (on page 11-328)
Trigger model (on page 5-24)
Scanning and triggering (on page 5-23)
scan.trigger.arm.stimulus
This attribute determines which event starts the scan.
Instrument reset
Recall setup
Scan reset
Create configuration script
Save setup
Usage
eventID = scan.trigger.arm.stimulus
scan.trigger.arm.stimulus = eventID
Trigger stimulus used for the channel action (arm layer); see Details
Details
This attribute selects which events cause the arm event detector to enter the detected state.
Set this attribute to 0 to start the scan without waiting for an event.