Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-413
Also see
trigger.timer[N].clear() (on page 11-412)
trigger.timer[N].delay (on page 11-413)
trigger.timer[N].reset() (on page 11-416)
trigger.timer[N].delay
This attribute sets and reads the timer delay.
Instrument reset
Recall setup
Trigger timer N reset
Create configuration script
Save setup
Usage
interval = trigger.timer[N].delay
trigger.timer[N].delay = interval
Delay interval in seconds (0.5 µs to 100 ks)
Trigger timer number (1 to 4)
Details
Once the timer is enabled, each time the timer is triggered, it uses this delay period.
Assigning a value to this attribute is equivalent to:
trigger.timer[N].delaylist = {interval}
This creates a delay list of one value.
Reading this attribute returns the delay interval that will be used the next time the timer is triggered.
Example
trigger.timer[1].delay = 50e-6
Set the trigger timer 1 to delay for 50 µs.
Also see
trigger.timer[N].reset() (on page 11-416)