EasyManuals Logo
Home>Keithley>Multimeter>3700 Series

Keithley 3700 Series User Manual

Keithley 3700 Series
936 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #596 background imageLoading...
Page #596 background image
Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-217
dmm.rel.enable
Enables or disables relative measurement control for the function selected by dmm.func.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Reset
DMM reset
Recall setup
Create configuration script
Save setup
0 (dmm.OFF)
Usage
value = dmm.rel.enable
dmm.rel.enable = value
value
The setting:
â–ª Enable: dmm.ON or 1
â–ª Disable: dmm.OFF or 0
Details
This attribute is not available if dmm.func is set to "continuity" or "nofunction". If you query
this attribute when either of these functions is selected, nil is returned.
When relative measurements are enabled, all subsequent measured readings are offset by the
relative offset specified by dmm.rel.level. Each returned measured relative reading will be the
result of the following calculation:
Relative reading = Actual measured reading – Relative offset value
If you change functions with dmm.func, the relative enable setting changes to the enable setting for
that function.
The relative enable setting is saved with the dmm.func function setting, so if you use another
function, then return to the previous function, the relative enable settings you set previously are
retained.
An error is generated if:
• dmm.func is set to "continuity" or "nofunction".
• If the value is out of range for the selected function.
Example
dmm.func = "accurrent"
dmm.rel.acquire()
dmm.rel.enable = dmm.ON
Enables the relative measurements for AC
current and uses the acquire command to set
the relative level attribute.
Also see
dmm.configure.recall() (on page 11-165)
dmm.configure.set() (on page 11-166)
dmm.func (on page 11-179)
dmm.rel.acquire() (on page 11-216)
dmm.rel.level (on page 11-218)

Table of Contents

Other manuals for Keithley 3700 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3700 Series and is the answer not in the manual?

Keithley 3700 Series Specifications

General IconGeneral
BrandKeithley
Model3700 Series
CategoryMultimeter
LanguageEnglish

Related product manuals