Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-431
tsplink.trigger[3].stimulus = scan.trigger.EVENT_CHANNEL_READY
Sets the trigger stimulus of the TSP-Link trigger line 3 event detector to scan the trigger channel ready event.
Also see
tsplink.trigger[N].assert() (on page 11-424)
tsplink.trigger[N].reset() (on page 11-429)
tsplink.trigger[N].wait()
This function waits for a trigger.
Usage
triggered = tsplink.trigger[N].wait(timeout)
Trigger detection indication; set to one of the following values:
true: A trigger is detected during the timeout period
false: A trigger is not detected during the timeout period
The trigger line (1 to 3)
The timeout value in seconds
Details
This function waits up to the timeout value for an input trigger. If one or more trigger events were
detected since the last time tsplink.trigger[N].wait() or tsplink.trigger[N].clear()
was called, this function returns immediately.
After waiting for a trigger with this function, the event detector is automatically reset and rearmed.
This is true regardless of the number of events detected.
Example
triggered = tsplink.trigger[3].wait(10)
print(triggered)
Waits up to 10 seconds for a trigger
on TSP-Link
®
line 3.
If false is returned, no trigger was
detected during the 10-second
timeout.
If true is returned, a trigger was
detected.
Also see
tsplink.trigger[N].clear() (on page 11-425)