Details
This command generates an error if the:
• Calibration is locked
• Step is out of sequence
• Step does not exist
• Step does not complete successfully
• Value passed is invalid for the step, out of range, or not needed
Example
For example of use, see DC volts adjustment (on page B-23), Resistance adjustment (on page B-26),
and DC current adjustment (on page B-27).
Also see
dmm.calibration.ac() (on page 11-155)
dmm.calibration.lock() (on page 11-156)
dmm.calibration.unlock() (on page 11-158)
dmm.calibration.lock()
This function locks calibration to prevent unintended changes.
Usage
dmm.calibration.lock()
Details
Use this command to lock an unlocked calibration. An error is generated if this command is issued
when calibration is already locked.
Once locked, you must unlock calibration before you can perform calibration again.
Save calibration data before locking. Calibration data will be lost if it is not saved before locking.