Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-306 3700AS-901-01 Rev. D/June 2018
Example
print(ptp.portstate)
Output (this output indicates that PTP is
disabled):
2.000000000e+00
Also see
ptp.enable (on page 11-305)
ptp.slavepreferred
This attribute describes whether you prefer to have the instrument be a subordinate clock or not.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Never
Nonvolatile memory
false (disabled)
Usage
value = ptp.slavepreferred
ptp.slavepreferred = value
value
0: Disabled
1: Enabled
Details
From the factory, this attribute is 1. After you set this attribute, it is saved in nonvolatile memory. That
setting is recalled the next time the instrument is powered up.
Example
ptp.slavepreferred = 1
print(ptp.slavepreferred)
Set the instrument to be a subordinate clock.
Check to see if the instrument is a
subordinate clock.
Output:
1
Also see
Not applicable

Table of Contents

Other manuals for Keithley 3700 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3700 Series and is the answer not in the manual?

Keithley 3700 Series Specifications

General IconGeneral
BrandKeithley
Model3700 Series
CategoryMultimeter
LanguageEnglish

Related product manuals