EasyManuals Logo
Home>Keithley>Multimeter>3700 Series

Keithley 3700 Series User Manual

Keithley 3700 Series
936 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #480 background imageLoading...
Page #480 background image
Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-101
channel.trigger[N].clear()
This function clears any pending events.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
channel.trigger[N].clear()
N
Number indicating the trigger line to clear (1 to 8)
Details
This function clears any pending events for the channel trigger specified by N.
Example
channel.trigger[1].clear()
Clears any pending events on channel
trigger 1.
Also see
channel.trigger[N].set() (on page 11-103)
channel.trigger[N].EVENT_ID
This constant indicates the trigger event generated by the channel trigger N.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
X = channel.trigger[N].EVENT_ID
X
The trigger event number
N
Number indicating the channel trigger event ID (1 to 8)
Example
scan.trigger.channel.stimulus = channel.trigger[1].EVENT_ID
Use channel trigger 1 events to pace the channel action of the scanning or set the trigger stimulus of the
channel event detector to channel trigger 1.
Also see
channel.trigger[N].set() (on page 11-103)

Table of Contents

Other manuals for Keithley 3700 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3700 Series and is the answer not in the manual?

Keithley 3700 Series Specifications

General IconGeneral
BrandKeithley
Model3700 Series
CategoryMultimeter
LanguageEnglish

Related product manuals