Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-368 3700AS-901-01 Rev. D/June 2018
Usage
pseudoCard = slot[slot].pseudocard
slot[slot].pseudocard = pseudoCard
pseudoCard
Set pseudocard to one of the following values:
â–ª slot.PSEUDO_NONE or 0 for no pseudocard selection
▪ 3720 for 3720 Dual 1 × 30 Multiplexer card simulation
▪ 3721 for 3721 Dual 1 × 20 Multiplex card simulation
▪ 3722 for 3722 Dual 1 × 48 Multiplexer card simulation
▪ 3723 for 3723 Dual 1 × 30 Reed Multiplexer card simulation
▪ 3724 for 3724 Dual 1 × 30 FET Multiplexer card simulation
▪ 3730 for 3730 6 × 16 High Density Matrix card simulation
▪ 3731 for 3731 6 × 16 High Speed Reed Relay Matrix card simulation
▪ 3732 or 37320 for 3732 Quad 4 × 28 Ultra-High Density Reed Relay Matrix card
simulation
▪ 37321 for 3732 Dual 4 × 56 Ultra-High Density Reed Relay Matrix card simulation
▪ 37322 for 3732 Single 4 × 112 Ultra-High Density Reed Relay Matrix card
simulation
▪ 37323 for 3732 Dual 8 × 28 Ultra-High Density Reed Relay Matrix card simulation
▪ 37324 for 3732 Single 16 × 28 Ultra-High Density Reed Relay Matrix card
simulation
â–ª 3740 for 3740 32-Channel Isolated Switch card simulation
â–ª 3750 for 3750 Multifunction I/O card
slot
Slot number (1 to 6)
Details
Pseudocards allow you to configure your system without having an actual plug-in card installed in
your system. You can perform open, close, and scan operations and configure your system with
pseudocards.
This command is only applicable to a slot that does not have a plug-in card or pseudocard installed. If
a pseudocard is presently assigned to the slot, you must set the slot to no pseudocard before
assigning the new pseudocard.
After assigning a pseudocard, you can use valid commands for the plug-in card for that slot.
Changing the pseudocard assignment from a pseudocard to no pseudocard invalidates scan lists that
include that slot.
If a module is installed in the slot, this command returns nil. If the slot is empty and no pseudocard
is installed, the return is 0.
Example
slot[6].pseudocard = 3720
Sets the pseudocard of slot 6 for 3720 plug-
in card simulation.
Also see
slot[slot].idn (on page 11-362)

Table of Contents

Other manuals for Keithley 3700 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3700 Series and is the answer not in the manual?

Keithley 3700 Series Specifications

General IconGeneral
BrandKeithley
Model3700 Series
CategoryMultimeter
LanguageEnglish

Related product manuals