108
Item Description
scanned information as the default item. For example, if the default item
is set to Patient ID and 2D and 1D match fail after barcode scanning,
the scanned data is automatically displayed in the Patient ID editing box
on the patient information page.
Move Up/Down Move up or down the selected item.
Add/Delete Add or delete the selected item. Only items that are set before delivery
can be added or deleted.
Restore default Restore the default values of items.
Worklist query
– If a worklist server is selected, the worklist server is queried based
on scanned information.
– If no worklist server is selected, a check is created on the patient
information page based on the scanned information.
Default query
item
– Select the default item for query on the worklist server.
For example, if Patient ID is selected, the worklist server is queried
based on the patient ID in the barcode.
– If no default query item is set, the worklist server is queried by
patient ID, sequence number, and patient information in order.
If both 1D and 2D default items are configured, the matching
priorities after barcode scanning is as follows: 2D, 1D, and
default items.
Import/Export Preconfigure barcodes by importing or exporting configuration files. For
details, Mindray Customer Service Department or sales representative.
"Advanced" parsing mode
Scan a barcode consisting certain information of the hospital, enter a regular expression
based on the barcode format, and tap Match to match the scanned barcode with the regular
expression. If they are matched successfully, the barcode information is displayed in multiple
items in the item list below.
– Default item: If no default item is set, the scanned information is directly input after 2D and
1D match failures. If a default item is set, the system considers scanned information as the
default item. For example, if the default item is set to Patient ID and 2D and 1D match fail
after barcode scanning, the system automatically fills the barcode data in the Patient ID
editing box on the patient information page.
Diagnostic Ultrasound System
Service Manual
3 Installation

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals