160
Table 5–7 Inspection of optional functions, peripherals and accessories(continued)
Item
Content
Method
3 Footswitch Check whether the footswitch implements the configured
functions according to the program.
4 DVD-R/W Check whether DVD-R/W is working properly (burning,
reading and ejecting).
5 Barcode scanner Check whether the scanner works normally and the output is
correct.
6 DICOM Check whether DICOM works normally. Send pictures and
other data to DICOM server for verification.
7
ECG Module Check the user's basic operations and verify the
implementation of functions of the ECG module.
55..11..33..44 Mechanical Safety Inspection
Mechanical safety inspection mainly focuses on the mechanical strength and mechanical function
of key components of the ultrasound system. The main inspection and evaluation methods
include visual inspection and operation inspection. If the inspection result fails to pass and proves
that the system is in an abnormal state, stop the system and take appropriate measures. The
inspection process is as follows:
Figure5–2 Mechanical safety inspection flowchart
The checklist for mechanical safety inspections is as follows:
Diagnostic Ultrasound System
Service Manual
5 Maintenance

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals