4. On the maintenance self-test main page, you can pre-configure the corresponding test items
by referring to the following table. Click Start to enter the maintenance self-test state after the
configuration is completed.
Button or Check
Box
Function
Open All Click this button to expand all the folded test items.
Close All Click this button to fold all the expanded test items.
Default Click this button to automatically select the default test items required
by the system, and the corresponding test block diagram will be
highlighted.
Select All Click this button to select all test items and all the test block diagrams
will be highlighted.
Clear All Click this button to deselect all test items and all the test block diagrams
are not highlighted.
Continue/Stop There are two options: Continue and Stop. If Continue is selected, it
means that the test will continue even if any test item fails during the
test. If Stop is selected, the test will be stopped once a test item fails
during the test.
Start Click this button to test the selected test items one by one, and
meanwhile the button will change to the Stop button. If the Stop button
is clicked during the test, the test will be terminated. After the test is
completed, the text on the Stop button will become Start again, and
can be clicked to start the next test.
Next Fail When the test is complete, click the Next Fail button, and the program
will automatically look for the next failed test item in the test information
list and display the test information of this item in the first row of the list.
If the last row of the list is reached, the query automatically starts from
the first row of the list.
Clear History Click this button to clear the test data in the test items and test
information and restore the system structure to the initial state.
Loops
Select this check box to start the loop test control. If this check box is
selected and the number of loops is entered in the right text box, the
test will be conducted according to the number of loops entered. If the
number of loops is less than 0, the loop tests will be conducted until the
user stops the test or the test is stopped when an error occurs.
Save Report Click this button to display the report export dialog box. Insert the
mobile storage device, select the corresponding logical drive letter, and
then click OK. The test data will be saved to the root directory of the
selected logical drive. If the mobile storage device is not inserted, the
button becomes unavailable.
Diagnostic Ultrasound System
Service Manual
5 Maintenance

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals