Page #169 background imageLoading...
Page #169 background image
Revision:1.0(2023-01-12)
170
Button or Check
Box
Function
Shut Down Click this button to close the program, and the machine shuts down
automatically. If the test report is not saved, the system will prompt you
whether to save the test report to the mobile storage medium before
shutdown.
System structure Clicking a module in the system structure will deselect all other modules
and highlight only the selected module. Continue to click other modules
to add or remove modules.
Note: In the test item area, the test items are sorted by board into multiple parent test items.
Each parent test item contains a plurality of sub-test items. Test items can be customized as
required. If the check box in front of the test item is selected (in
state), the test will be put
into the test sequence; otherwise, if it is cleared (in
state), the test will not be put into the
test sequence. When a parent test item is selected or deselected, the child test items
included in the parent test item are selected or deselected by default.
The user can see production, maintenance and user self-test pages from the screen. The
production and maintenance self-test pages can be divided into five areas.
Figure5–4 Maintenance self-test page
5. Self-test status display
Diagnostic Ultrasound System
Service Manual
5 Maintenance

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals