Revision:1.0(2023-01-12)
172
Figure5–6 Display of test results in the test item list
Click a test item in the list, and the program will automatically index the detailed test results
corresponding to the test item, and display the test results in detail, as shown in the figure
below.
Figure5–7 Display of test results in the message list
All test items are automatically numbered in the format of ZXXYY, where XX indicates the
sequence number of the level-1 test item where this test item belongs, and YY indicates the
sequence number of the level-2 test item. For example, "Parameter board I2C communication
test" is numbered Z0801, where 08 indicates that "Main display" is the eighth test item under
Diagnostic Ultrasound System
Service Manual