Removal and Installation Procedures
Front-Panel Trim,
Menu and Power
Buttons,
EMI Gasket
6·18
Rear Cover
1.
Position the oscilloscope as indicated in Figure
6-7.
2.
Use the screwdriver to remove the four
T-15
Torx screws that secure the
rear cover to the
oscilloscope.
3.
Lift off the rear cover.
Do
these steps
in
reverse order to install the rear
cover.
Cabinet
1 . Remove the rear cover.
2.
Remove the Single
T-15
Torx screw on the left side of the oscilloscope.
3.
Grasp the two handle hubs and pull them outward
as
if to rotate the
handle.
4.
While holding the handle hubs out, lift the cabinet up and slide
it
off the
oscilloscope.
Take
care not to bind or snag the cabinet on the internal
cabling
as
you remove
it.
Replacing the Front Feet
If you are replacing a front foot, use a pair of slip-jaw pliers to firmly grip the
foot, then pull with a turning motion
to
remove. Do this step in reverse order
to
replace the foot.
Use the following procedures to remove and
install front-panel trim, buttons,
and
EMI
gaskets.
Additional procedures required - These procedures require that the
following procedures be performed first:
• Rear Cover Removal
•
Cabinet Removal
Removing the Front Trim
1.
Slide a flat-bladed spudger between the chassis and the plastic front
trim, near one of the four
flex locks. See Figure 6-8.
2.
Gently pry the front trim up and slightly forward to disengage the flex
lock.
Do this until all four flex locks are disengaged.
3.
Pull the front trim forward and off the instrument.
Maintenance
•
•
•