This procedure assumes that
there
is
front panel control.
Measure the power supply
voltages
on
J67 on the A 1
Analog board.
Pin
12
+15V
Pin
10 +B.6V
Pin
9
-B.6V
Pin
6 +5.2V
Pin
1
-5.2V
Press the CH1 button
and rotate the
VOLTS/DIV
knob several times
in
each direction. Repeat
for remaining
channel(s).
Listen for clicks between
some of the
voltage
ranges.
Turn
the power
off.
Remove
the
cable at J67 on the A 1
Analog board and check
pins
12,
10,
9,
6,
and 1 for
shorts to ground.
Replace the
A1
Analog board.
Troubleshooting
Perform the
A3
Display Driver
Board Troubleshooting
Procedure.
Perform both to
isolate
the problem.
Figure 6-22: A 1 Analog Board Troubleshooting Procedure
TAS 475 and TAS 485 Service Manual