Page #146 background image
Troubleshooting
Put the oscilloscope in
XV
mode. Select CH2 and turn
the readout intensity to
minimum. Measure the
voltage across the delay
line
(J68) with a DMM.
Select
CH1
and
measure the voltage
across pins
16
and
17
of J67 with a DMM or
oscilloscope.
Rotate the Intensity knob
and measure the voltage at
pin
19 of J67 with a
DMM
or oscilloscope.
Note:
In
XV
mode, a single
dot should appear near
the center of the
CRT.
Reduce the intensity until
the spot
is
dim so that the
CRT
phosphor will not be
damaged.
No
No
No
Perform the
AS
CPU
Board Troubleshooting
Procedure.
Replace the
AS
CPU
board.
Figure 6-22: A 1 Analog Board Troubleshooting Procedure (Cont.)
6-44
Maintenance
Scan by Zenith