Warranted Characteristics
Table 1-11: Warranted Characteristics - Time Base System (Cont.)
Name
TAS 475 Time Base Accuracy,
Magnify
On
TAS 485 Time Base Accuracy,
Magnify
On
Horizontal Position Control Range
Variable Control Range
1-12
Description
The limits are
as
follows:
Conditions
Time
Measurement Accuracy
-10°
C to + 55 ° C
Excluding
5 ns/div and
±4%.
2 ns/div
5 ns/div and 2 ns/div
±3%.
-10°
Cto
+ 15°
Cand
+
35
° C to +
55
° C
5 ns/div and 2 ns/div
±5%.
Sweep accuracy applies over the center eight divisions. Excludes
the greater of either the first
Y4
division or
25
ns from the sweep start
of
the magnified sweep and anything beyond the 100th magnified
division.
The limits are
as
follows:
Conditions
Time
Measurement Accuracy
-10°
Cto+55°
C
Excluding
2 ns/div and
±4%.
1 ns/div
2 ns/div and 1 ns/div
±3%.
-10°
C to + 15° C and
+
35
° C to +
55
° C
2 ns/div and 1 ns/div
±7%.
Sweep accuracy applies over the center eight divisions. Excludes
the greater of either the first
Y4
division or
25
ns
from the sweep start
of the magnified sweep and anything beyond the 1
DOth
magnified
division.
The position
control is able to move the start of the sweep to the
right of the center
vertical graticule and able to move a time mark
corresponding to the tenth division
on
an
unmagnified sweep to the
left of the graticule center.
Continuously variable between calibrated seconds/division settings.
Extends both the Main and
Delayed sweep seconds/division set-
tings by a factor of 2.5 times.
Specifications
•
•
•