THRESHOLD: Configures the threshold range. Related command: dmm.threshold (on page 11-232).
UNITS: Configures the units for voltage and temperature measurements. Related command:
dmm.units (on page 11-234).
LIMIT key configuration
Pressing the CONFIG key and then the LIMIT key opens the LIMIT menu. Select LIMIT 1 or LIMIT 2
to open the LIMIT 1 or LIMIT 2 menu.
These menus contain the following items:
• ENABLE: Enables or disables limit testing. Related command: dmm.limit[Y].enable (on page 11-
184).
• CLEAR: Clears the test results of the limit. Related command: dmm.limit[Y].clear() (on page 11-
184).
• AUTOCLEAR: Indicates if the limit should be cleared automatically or not. Related command:
dmm.limit[Y].autoclear (on page 11-183).
• LOWVAL: Sets the low limit value. Related command: dmm.limit[Y].low.value (on page 11-193).
• LOWFAIL: Queries for the low test results of the limit. Related command: dmm.limit[Y].low.fail
(on page 11-191).
• HIGHVAL: Sets the high limit value. Related command: dmm.limit[Y].high.value (on page 11-
189).
• HIGHFAIL: Queries for the high test results of limit. Related command: dmm.limit[Y].high.fail (on
page 11-187).
REL key configuration
Press the CONFIG key and then the REL key to open the RELATIVE OFFSET menu.
The RELATIVE OFFSET menu contains the following menu items:
• ACQUIRE: Acquires an internal measurement to store as the REL level value. Related
command: dmm.rel.acquire() (on page 11-216).
• ENABLE: Enables or disables relative measurement control for the DMM. Related command:
dmm.rel.enable (on page 11-217).
• LEVEL: Sets a specific offset value to use for relative measurements for the DMM. Related
command: dmm.rel.level (on page 11-218).
FILTER key configuration
Press the CONFIG key and then the FILTER key to open the FILTER menu.
The FILTER menu contains the following menu items:
• ENABLE: Enables or disables filtered measurements for the selected DMM function. Related
command: dmm.filter.enable (on page 11-175).
• COUNT: Indicates the filter count setting for the selected DMM function. Related command:
dmm.filter.count (on page 11-174).
• TYPE: Indicates the filter averaging type for the DMM measurements on the selected DMM
functions (MOVING or REPEAT). Related command: dmm.filter.type (on page 11-176).
• WINDOW: Indicates the filter window for the DMM measurements (0 to 10% in 0.1% increments).
Related command: dmm.filter.window (on page 11-177).