Revision:1.0(2023-01-12)
138
Procedure Acceptance criteria
Tap ImageMenu>A. Power for adjustment. The A. Power value is displayed in real time
in the upper part of the screen.
U/D Flip
Tap ImageMenu>U/D Flip for adjustment.
It can rotate images upward or downward.
The
mark indicates the default
orientation of an image in the upper left
corner.
Rotation
Tap ImageMenu>Rotation for adjustment.
The system supports rotation by an angle of
0°, 90°, 180°, or 270°. When the image is
rotated in an angle of 90° or 270°, the depth
scale is displayed on the top of the screen.
The
mark indicates the default
orientation of an image in the upper part.
H Scale
Tap ImageMenu>H Scale for adjustment.
Tap H Scale in the image menu to display or
hide the horizontal scale (width mark).
Colorize
Tap ImageMenu>Colorize for adjustment.
Echo Boost
Tap ImageMenu>Echo Boost to turn on or off
Echo Boost.
Use phased probe to activate the function in
cardiac mode.
M mode
In M mode image scanning, the image parameter area on the left side of the screen displays the
real-time parameter values as follows:
Parameter F D G V
Description Frequency Depth M gain M speed
Parameters that can be adjusted to optimize the M mode image are indicated in the following.
Adjustment method Parameter
Adjustment in the image area Gain, Depth, iTouch, TGC, and Image quality
Adjustment in the image menu Display Format, Speed, Dynamic Range, A.
Power (%), Dual, iClear, M Soften, Gray Map,
Tint Map, and Colorize
Color mode
Diagnostic Ultrasound System
Service Manual
4 Function and Performance
Check