Clarus 600/680 User’s Guide
4-29
TCD Setup
Setting up a TCD requires that you equalize carrier gas flows through both channels,
followed by balancing the current through the two filaments. The TCD is
nondestructive and can be used in series with other detectors. The TCD is a
universal detector that is concentration sensitive. The TCD sensitivity depends on
the filament current, component concentrations, and the difference in thermal
conductivity between the carrier gas and the component.
NOTE: To prevent TCD filament damage, shut off the filament current before
disconnecting columns or turning off gas flows.
To set up a TCD, read the TCD setup summary, then follow the steps.
TCD Setup Summary
To set up a TCD:
• Ensure that a column is installed.
• Turn on the oven and detector heaters.
• Equalize the makeup and reference gas flows.
• Enter the TCD bridge current.
• Balance the bridge
Turn on the Oven and Detector Heaters
1. Close the oven door. This turns on the oven heater.
2. From the System Status screen touch the TCD button
.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer CLARUS 680 GC and is the answer not in the manual?

PerkinElmer CLARUS 680 GC Specifications

General IconGeneral
BrandPerkinElmer
ModelCLARUS 680 GC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals