Setting up a Typical Analysis
7-4
System Status Screen
Everything starts from the system status screen.
1. Touch the Oven button
to display the oven information.
2. The oven information is now visible. To program the oven temperatures
touch the temperature field and it will turn black (highlighted). Then use
the up and down arrow or keypad
buttons to set to the
required temperature for your application. Repeat the same procedure for
the time and rate.
As shown in the example below this analysis has been set to have an initial
temperature of 75 ºC and it is held at this temperature for 1 minute then
goes up 10 degrees/minute.