Clarus 600/680 User’s Guide
3-25
Setting Carrier Gas Flow using PPC
To view or set the carrier flow:
1. From the System Status screen touch the Injector button (for example, Pkd or
Cap) corresponding to the injector you wish to set the carrier gas flow. The tab
for the selected injector appears:
In this example, the Clarus GC is configured with the PSS capillary mode on for
flow. If the option for capillary control has been selected during configuration,
you can program the carrier gas according to one of three values: pressure,
velocity, or flow. The carrier gas field on the injector tab screen displays the set
point value. The label below the field displays the current step. The actual values
for all the available configuration options are displayed to the left as read only text
on the screen.
2. Touch the program button
for Carrier Gas.