Using the Active Method
3-18
Inlet Temperature Programming Mode
The inlet temperature programming mode is used when the POC or PSS injector is
configured as Inlet. In this mode, the injector follows the injector temperature program
that you entered in the method.
The initial default values are: Initial (Temp 1) is 50 °C,
Injection Time is 999 min, and Inj Rate is 0. With non-
programmed injectors, an entry of 0 turns the heated zone
off. With programmed injectors, a temperature of 0 is
possible. Therefore, when 0 is entered for the initial
injector setpoint, the following screen appears:
1. From the selected injector tab screen make sure that the injector heater is off.
2. Set the initial temperature to 50
°C.