Using the Tools Menu
6-18
Oven
Use the oven configuration screen to set up oven parameters. This example
shows a temperature limit of 300 °C and an Equilibration rate of 1.0 mL/min.
Calibration
When the Calibration Offset field is selected it turns black.
The default value is 0. The Offset range is -10 to + 10 C. Enter the Offset
temperature and allow the oven temperature to equilibrate for at least 30 minutes.
The displayed temperature will approach 100.00
°C, or the oven temperature you
entered, and the thermometer readout will approach a value close to 100.00
°C.
Coolant
If you have an optional coolant select the type installed, its target temperature
and a timeout time.