Clarus 600/680 User’s Guide
4-57
Setting Up the FPD
To set up a FPD using PPC modules, follow the steps below.
FPD Setup Summary
To set up a FPD:
• Configure the linearizer and signal filter.
• Set the FPD temperature.
• Set the hydrogen flow and air flow for PPC
• Turn on the photomultiplier tube.
• Ignite the flame
Configure the Linearizer and Signal Filter
The signal for sulfur is a function of the square of the sample quantity injected.
That is, if twice as much sample is injected, the response is quadrupled. When the
linearizer is turned on, the software calculates the square root of the signal and sends
that value to the integrator or recorder.
NOTE: If Linearization is set to "ON," the Autozero function must be checked off on the
FPD tab screen.
NOTE: If Linearization is set to "ON," set the Clarus GC integrator attenuation to x64. If
Linearization is set to "OFF," set the Clarus GC integrator attenuation to x8.
1. From the System Status screen touch the FPD button.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer CLARUS 680 GC and is the answer not in the manual?

PerkinElmer CLARUS 680 GC Specifications

General IconGeneral
BrandPerkinElmer
ModelCLARUS 680 GC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals