Technical data Sound level
106
4522 207 12671/ * 2021-06-17
Pathology Scanner SG20 / SG60 / SG300
NOTICE
Only qualified and authorized service engineers are allowed to replace the fuses.
Sound level
Definition Specification
Noise emission during operation < 60 dBA (A weighted, distance 1 m)
Tab. 54: Sound level
Electromagnetic compatibility
Electromagnetic emissions
The scanner complies with the EMC classification mentioned below.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions Group 1 The scanner RF emissions are very low and are not
likely to cause any interference in nearby electron-
ic equipment.
Class A
Tab. 55: Electromagnetic emissions
This product has been tested and found to comply with the limits for a Class A digital device
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated in a commercial
environment. This product generates, uses, and can radiate radio frequency energy and, if not
installed and used in accordance with the manufacturer’s instruction manual, may cause
harmful interference with radio communications. Operation of this product in a residential area
is likely to cause harmful interference, in which case you will be required to correct the
interference at your own expense.
This equipment complies with the emission and immunity requirements described in the IEC
61326-2-6.
WARNING
This equipment has been designed and tested to CISPR 11 Class A. In a domestic
environment it may cause radio interference, in which case, you may need to take measures
to mitigate the interference.
WARNING
Do not use this device in close proximity to sources of strong electromagnetic radiation (e.g.
unshielded intentional RF sources), as these can interfere with the proper operation.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips Pathology Scanner SG20 and is the answer not in the manual?

Philips Pathology Scanner SG20 Specifications

General IconGeneral
BrandPhilips
ModelPathology Scanner SG20
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals