Administrator tasks Manual calibration
44
4522 207 12671/ * 2021-06-17
Pathology Scanner SG20 / SG60 / SG300
Administrator tasks
Manual calibration
Administrator authorization rights are necessary to manual calibrate the scanner. The
administrator users can start a manual calibration, if desired. Note that the scanner will
automatically calibrate and that manual calibration is not necessary.
Open the Settings screen.
Select Maintenance.
CALIBRATE
Maintenance
Settings
Scanner Calibration
Fig. 24: Settings screen Maintenance tab
Select Calibrate to initiate a calibration manually.
Select Calibrate in the pop-up window to confirm the calibration request.
The scanner will perform all required calibrations, this can take several minutes.
A calibration status message will be displayed.
NOTICE
A calibration request cannot be undone.
If the Calibrate button is selected again during calibration, this will only re-start the calibration
and will NOT result in stopping the calibration process.
On the user interface, select Unlock door and open the store door.
Close the store door.
Closing the doors confirms that the scanning process can continue.