EasyManuals Logo

Philips Pathology Scanner SG20 User Manual

Philips Pathology Scanner SG20
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #50 background imageLoading...
Page #50 background image
Cleaning Cleaning instructions
50
4522 207 12671/ * 2021-06-17
Pathology Scanner SG20 / SG60 / SG300
Frequency What to clean Material to use
Exterior of the scanner Lint free cloth, either dry or damp with cleaning liquids (i.e. isoprop-
yl alcohol)
When the error
'Calibration
failed' occur-
red.
Clean the following items
as described in daily clean-
ing:
Stage
Light source lens
Diffuser
Swabs with a microfiber fabric over foam tip, either dry or damp
with cleaning liquids (i.e. isopropyl alcohol)
Tab. 14: Cleaning overview
Cleaning instructions
Daily cleaning
1. Wait until all slides are processed and returned to the store.
2. Unload all racks from the store.
3. Use the power button on the front side of the scanner to shut down the scanner.
4. Select Yes to confirm.
5. Wait for the blue LED in the power button to turn off.
6. Use the power switch (I) on the
rear side of the scanner to stop the
power supply.
I
7. Push the pin of the store door lock
on the bottom of the scanner.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips Pathology Scanner SG20 and is the answer not in the manual?

Philips Pathology Scanner SG20 Specifications

General IconGeneral
BrandPhilips
ModelPathology Scanner SG20
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals