*353074 – Average density 1000,0
*351175 – Factor 1 0,0
*351176 – Factor 2 0,0
*351177 – Factor 3 0,0
*351178 – Min. temp. 0
*351179 – Max. temp. 0
*35309 – Float correction 0
9.7.5.1.12. Temp. Compensation – 35117
Select the Temp. Compensation
menu with <7>. Specific temperature
compensation values can be entered
and defined here. Also see "Table of
all used abbreviations".
With <5> Factor 1:
Linear interpolation (LIN):
V
0
= V
T
× (1 – k
0E

• Factor 1 (k
0E
) defined by the PTB for various products.
• No distinction between summer and winter.
Factors and temperature range can be adjusted in Setup (W&M protected).
With <5> - <7> Factor 1-3:
Polynomial level 3 (POL):
V
0
= V
T
× {1+ [(A
1

2

2
) + (A
3

3
)]}
• Factor 1-3 (A
1
– A
3
) defined for various products by the PTB.
• No distinction between summer and winter.
Factors and temperature range can be adjusted in Setup (W&M protected).
With <8> as the min. product temperature for LIN & POL.
With <9> as the max. product temperature for LIN & POL.
MultiLevel Instruction Manual MENU structure

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TechnipFMC Sening MultiLevel and is the answer not in the manual?

TechnipFMC Sening MultiLevel Specifications

General IconGeneral
BrandTechnipFMC
ModelSening MultiLevel
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals