TBN581 (ORP/pION) TRANSMITTER
CALIBRATION
L - 8
Bench Calibration (mV)
This procedure applies to TBN581 Transmitters configured for either
ANALOG or DIGITAL. Any difference between the two are noted in
the Comments column.
This value shows the offset change over time for a
particular sensor. Value resets to 0.00 mV if installing
a new sensor.
Calibration complete.
Key Display Comments
ENTER
CUMULATIVE
OFFSET CHANGE
nnn pH
HIT ENTER
ENTER
ID TAGNAME
READY
Key Display Comments
This operation will cause a change in output not cor-
responding to the input. Be sure the control loop is in
manual.
Select YES.
Select BENCH CAL.
Select mV SOURCE.
Low default value: -1000 mV.
High default value: +1000 mV.
Enter low mV value and press . Enter high
mV value.
CALI-
BRATE
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
→ YES
ENTER
CALIBRATION
PROCESS CAL
→ BENCH CAL
D-TO-A ADJUST
ENTER
INPUT SOURCE:
SAME SENSOR
NEW SENSOR
→ mV SOURCE
ENTER
ENTER LO mV CAL
nn.nn mV
ENTER HI mV CAL
nn.nn mV
ENTER