EasyManuals Logo

ABB STT04 User Manual

ABB STT04
285 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #214 background imageLoading...
Page #214 background image
TB82 TRANSMITTER
CALIBRATION
N - 8
Temperature Calibration
This procedure describes how to perform a temperature calibration.
Key Display Comments
This operation will cause a change in output not cor-
responding to the input. Selecting YES will hold the
output during calibration.
Select YES.
Select TEMP. CAL.
Enter the temperature. Default temperature is 25°C.
CALI-
BRATE
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
→ YES
ENTER
nnnn CAL
→ TEMP. CAL
EDIT CAL
RESET CAL
ENTER
ENTER
TEMPERATURE:
nn.nn
O
C
ENTER
ID TAGNAME
READY

Table of Contents