7-6 ApexPro™ Telemetry 2001989-203C
Parts lists, drawings and replacement
Drive filter
1. Turn the latch securing the front service door on the front panel of the ATS and
open the door.
2. Using a phillips screwdriver, remove the two screws securing the drive filter
cover.
012A
3. Remove the cover and filter.
4. Reverse the steps in this procedure to install the drive filter.
Chassis cover
1. Complete the preparation procedures. See Field replacement preparation on page
7-3.
2. Using a phillips screwdriver, remove the 4 screws securing the front side of the
cover.
041A
3. Using a phillips screwdriver, remove the 2 screws securing the back side of the
cover.
042A
4. Slide the cover backwards until the lip on the front of the cover clears the groove
in the chassis. Lift and remove the cover.
5. Reverse the steps in this procedure to install the chassis cover.