7-22 ApexPro™ Telemetry 2001989-203C
Parts lists, drawings and replacement
NOTE
The rebuild procedure takes approximately 45 minutes.
CD ROM drive
1. Complete the preparation procedures. See Field replacement preparation on page
7-3.
2. Remove the chassis cover. See Chassis cover on page 7-6.
3. Remove the drive bay. See Drive bay on page 7-16.
4. Using a phillips screwdriver, remove the 4 anchor screws securing the CD/hard
drive carrier in the drive bay on both sides.
018A
5. Slide the CD/hard drive carrier out of the front of the drive bay.
160A
6. Using a phillips screwdriver, remove the 2 screws securing the left side of the CD
ROM drive into the CD/hard drive carrier.
063A
7. Using a phillips screwdriver, remove the 2 screws securing the right side of the
CD ROM drive into the CD/hard drive carrier.