D-2 ApexPro™ Telemetry 2001989-203C
Electromagnetic compatibility
Electromagnetic compatibility (EMC)
Changes or modifications to this system not expressly approved by GE could cause
EMC issues with this or other equipment. This system is designed and tested to
comply with applicable regulation regarding EMC and needs to be installed and put
into service according to the EMC information stated in this appendix.
CAUTION
Use of portable phones or other radio frequency (RF) emitting
equipment near the system may cause unexpected or adverse
operation.
CAUTION
The equipment or system should not be used adjacent to, or stacked
with, other equipment. If adjacent or stacked use is necessary, the
equipment or system should be tested to verify normal operation in
the configuration in which it is being used.
Guidance and manufacturer’s declaration
Electromagnetic emissions
The ATS is intended for use in the electromagnetic environment specified below. It is
the responsibility of the customer or user to assure that the ATS is used in such an
environment.
Emissions test Compliance Electromagnetic environment – guidance
RF Emissions
EN 55022
Group 1 The device uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF Emissions
EN 55022
Class A
Harmonic Emissions
EN 61000-3-2
Class D
Mains power quality should be that of a typical commercial and/or hospital
environment.
Voltage Fluctuations/
Flicker Emissions
EN 61000-3-3
Complies