A-4 ApexPro™ Telemetry 2001989-203C
Technical specifications
Environmental specifications
Physical specifications
Certification
IEC/EN/UL 60950-1
CAN/CSA C22.2 No.950
IEC/EN 60601-1-4
EN 55022 (Class A)
EN 55024
EN 61000-3-2
EN 61000-3-3
CE marked to the Medical Device Directive 93/42/EEC
Backplane CPU/ 4-PCI.
Power supply 300 watt ATX dual-redundant.
Power requirements Input Range: 100-240 VAC Full Range.
Frequency: 50-60 Hz.
Input Current: 4.0 A.
Power consumption 300-320 watts.
Cooling 42 - 57 CFM Forced air with front accessible filter.
Operating environment
specifications
Temperature: 10 to 35 C (50 to 95 F).
Altitude: 0 to 3,048 meters (0 to 10,000 ft).
Relative Humidity: 15 to 80% (non-condensing).
Non-operating storage
environment specifications
Temperature: -23 to +49 C (-10 to 120 F).
Relative Humidity: 10 to 90% (non-condensing).
Vibration (operating) 5 to 500Hz, 0.5 grams.
Random vibration (non-operating) 5 to 20Hz, 0.001 to 0.01 G2 per Hz, 20 to 500Hz, 0.01
G2 per.
Shock (operating) 2.0 G with 11 mSec duration, 1/2 sine wave.
Acoustic noise Less than 52 dB sound pressure at 5 to 28 C (41 to
82 F).
Height 8.8 cm (3.5 in)
Depth 45 cm (17.7 in)
Depth w/handles 49.5 cm (19.5 in)
Width 48.2 cm (19 in)
Weight 15.8 kg (35 lbs.)

Table of Contents

Other manuals for GE ApexPro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ApexPro and is the answer not in the manual?

GE ApexPro Specifications

General IconGeneral
BrandGE
ModelApexPro
CategoryMedical Equipment
LanguageEnglish

Related product manuals