5-20 ApexPro™ Telemetry 2001989-203C
Configuration
127A
7. In the Master/Slave field, select Master to configure the ATS as the Device
Master, or select Slave to configure the ATS to act as a slave.
NOTE
Only one ATS can be the master; all other ATSs must be designated as
slaves.
Set the ApexPro FH Telemetry System network name and IP
address
WARNING
LOSS OF MONITORING/DATA — Be extremely careful not to
assign the same IP address to two different devices. Duplicate IP
addresses will cause erratic system communication and data loss.
If an ATS with a duplicate IP address is connected to the Unity
network, the Windows OS will return the IP address to 0.0.0.0 and
the original/default IP is lost and will reset. To resolve the
duplicate IP address issue, change the IP address in Webmin, and
reboot the ATS after changing the IP address.
Because this situation is most likely to occur if multiple devices are
ghosted simultaneously, complete the ghost procedure offline for
one ATS before beginning the procedure on another ATS.
WARNING
If a USB disk is plugged into the ATS, you need to remove it prior to
rebooting. Otherwise, the system software may not restart.

Table of Contents

Other manuals for GE ApexPro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ApexPro and is the answer not in the manual?

GE ApexPro Specifications

General IconGeneral
BrandGE
ModelApexPro
CategoryMedical Equipment
LanguageEnglish

Related product manuals