EasyManuals Logo
Home>Mindray>Medical Equipment>TE5

Mindray TE5 User Manual

Mindray TE5
323 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #215 background imageLoading...
Page #215 background image
Probes and Biopsy 13-15
Do not soak the probe connector or the cable near it into water or any solution.
Follow local regulations when selecting and using the sterilization solution.
3. Rinse the probe with plenty of sterile water (about 2 gallons) for at least 1 minute to remove all
chemical residues on it. Or, follow the rinsing method recommended by the sterilization
solution manufacturer to rinse the probe.
4. Wipe off the water on the probe with sterile cloth or gauze after rinsing it. Do not dry the probe
by heating.
Immerse the probe in the solution (take 7LT4s as an illustration)
NOTE:
Repeated disinfection may degrade the performance and safety of the probe.
13.1.6 Storage and Transportation
When all examinations for the session have been completed, confirm that the probe is in good
condition. After disinfecting the probe, confirm that the probe is still in good condition and stored in
a suitable place.
1. To prevent the probe from being damaged, DO NOT store it where it may be exposed to:
Direct sunlight or X-rays
Sudden changes in temperature
Dust
Excessive vibration
Heat generators
2. Store and transport the probe (L12-4s, L14-6Ns, V11-3Ws, SC6-1s, SP5-1s, 7L4s, L9-3s, L11-
3VNs, L12-3RCs, L14-5Ws and C11-3s) under the following ambient conditions:
Ambient temperature: -20°C to 55°C
Relative humidity: 20% to 95% (no condensation)
Atmospheric pressure: 700 hPa to 1060 hPa
Store and transport the probe (C5-2s, L7-3s, L14-6s, P4-2s, P10-4s, P7-3s, 6CV1s, C5-1s and
7LT4s) under the following ambient conditions:
Ambient temperature: -20°C to 55°C
Relative humidity: 30% to 95% (no condensation)
Atmospheric pressure: 700 hPa to 1060 hPa
Store and transport the P7-3Ts probe under the following ambient conditions:
Ambient temperature: -10°C to 45°C
Relative humidity: 30% to 90% (no condensation)
Atmospheric pressure: 700 hPa to 1060 hPa
Store and transport the L14-5sp probe under the following ambient conditions:
Ambient temperature: 0°C to 60°C
Relative humidity: 30% to 95% (no condensation)
Atmospheric pressure: 700 hPa to 1060 hPa
Store and transport the L20-5s probe under the following ambient conditions:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TE5 and is the answer not in the manual?

Mindray TE5 Specifications

General IconGeneral
BrandMindray
ModelTE5
CategoryMedical Equipment
LanguageEnglish

Related product manuals