System Maintenance 17-5
Compatible Disinfectants
The following table lists compatible disinfectants.
Advanced Ultrasound Solutions Inc.
Protex
TM
Disinfectant Spray
Sani-Cloth® Plus germicidal
disposable cloth
Schulke mikrozid
®
Sensitive
Wipes
Disinfection Procedures
Clean the system in power off status; or in power on status, tap in the top-right
corner of the screen and select to lock the system for 10 seconds.
The ultrasound system has passed puncture test and leakage current test, and thus
can be disinfected on power-on status.
After cleaning, disinfect the main unit as follows.
1. Wear medical gloves to prevent infection.
2. Disinfect the main unit with disinfectant wipe, spray or solution. Follow the disinfectant
manufacturer’s recommended contact time and method.
NOTE: Do not disinfect handle, or any visible sockets or interfaces (such as probe sockets,
ventilation holes, dust-proof cover, loudspeaker, sockets or interfaces in the IO panel and
power supply panel). See the following figure.