EasyManuals Logo
Home>Siemens>Measuring Instruments>SC 7000

Siemens SC 7000 Service Manual

Siemens SC 7000
160 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #91 background imageLoading...
Page #91 background image
ASK-T898-03-7600 Siemens Medical Systems, EM-PCS Danvers 75
7k9kXLSM.c5.CD_ROM.fm/04-99/kaupp
NOT A CONTROLLED DOCUMENT
Chapter 5: Troubleshooting
1Introduction
Troubleshooting SC 7000 and SC 9000XL monitors is either a two- step or
three-step process.
Step 1) Determine whether the trouble is external, such as a
malfunctioning module, Pod, connector, or cable, or internal.
Step 2) Do either a or b as appropriate.
a If external, recalibrate or replace malfunctioning item.
b If internal, go on to step 3.
Step 3) Usual remedial action is to replace malfunctioning component
or subassembly. See “Appendix A: Replacement Parts” for
complete listing of replaceable system components and
subassemblies. After replacing an internal subassembly,
functionally verify all monitor functions before returning monitor
to clinical service.
SC 7000 and SC 9000XL monitoring system devices consist of the Base
Unit, and peripheral devices including the Docking Station, CPS, IDS w/
Power Supply, R50 Recorder, etCO
2
Module, MultiMed or MultiMed 12
Pod, HEMO or HemoMed Pod, and Remote Display.
2 Recommended Tools and Test Equipment
Use the specified, recommended tools and test equipment, or a known
equivalent, when troubleshooting SC 7000 / SC 9000XL monitors and
peripheral equipment. Substitutions are approved only if an equivalent is
listed. Use of other test equipment and/or accessories could result in
inconclusive tests or damage to system components.
Table 5-1 Recommended Tools and Test Equipment
Tools & Test Equipment Description
Digital Multimeter (DMM), 4.5 digit Fluke, model 8050A (or equiv.)
Patient ECG/RESP, Temp, IBP simulator DNI Medsim 300B or equivalent
Leads
:
Three-lead grabber set, or IEC color code 1
IEC color code 2
Art. No. 33 75 230 E530U
Art. No. 33 75 248 E530U
Five-lead grabber set IEC color code 1
IEC color code 2
Art. No. 33 75 255 E530U
Art. No. 33 75 263 E530U
Patient SpO
2
simulator, or Nellcor PT2500 or equivalent
Reusable SpO
2
sensor:
Durasensor, Adult Art. No. 45 34 475 EH50U
NBP simulator (calibrated) or DNI CuffLink
Mercury manometer with hand bulb, or Baumanometer, 0-300 mmHg
Electronic pressure indicator with handbulb Veri-Cal Pressure Transducer Tester, or
equivalent
NBP Calibration Assembly Art. No. 28 77 855 EE54U
NBP connection hose 3.7 m Art. No. 12 75 275 EH40U