Climatic conditions
Ambient humidity (rel. humidity)
during storage, transport or opera‐
tion
<90%
Permissible ambient temperature
and humidity
Technical specications of analyzer modules (Page163)
13.3 Technical specications of analyzer modules
13.3.1 OXYMAT 7 technical specications
Table 13-7 OXYMAT 7: General technical specications
General OXYMAT 7 OXYMAT 7 high-tempera‐
ture
1)
Information on the smallest span See module nameplate or device certicate
Max. power consumption Warm-up phase ≤60W ≤120W
Weight (only module) < 5.5kg < 7.5kg
Electrical safety In accordance with DINEN61010-1
Max. EMC error during operation with one and/or two
OXYMAT 7 / ULTRAMAT 7 units
1.0% according to EN
61326-1 (2013)
1)
The high-temperature version of the analyzer module is not intended for installation in the rack-mounted device.
Table 13-8 OXYMAT 7: Measuring ranges
Measuring ranges OXYMAT 7 OXYMAT 7 high-temperature
1)
Number of measuring ranges Max. 4; parameters can be as‐
signed freely
Max. 4; parameters can be as‐
signed freely
Parameters can be as‐
signed in the measur‐
ing ranges
Smallest possible measuring
span
0.5% O
2
2% O
2
Largest possible span 100% O
2
100% O
2
1)
The high-temperature version of the analyzer module is not intended for installation in the rack-mounted device.
Note
Temperature change for high-temperature OXYMAT 7
If you change the temperature range, the device has to be re-calibrated. Detailed information
can be found in the operating manuals, see "Bibliography" (Page179).
Technical specications
13.3Technical specications of analyzer modules
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB 163