5.4.2 Notes on applications planning
Note
Usability of the device
Before you use the device, use the technical specications to check that your device is suitable
for the measuring task.
Note
Vibrations at the installation site disturb the measured signal
When you use OXYMAT7 and/or ULTRAMAT7 analyzer modules, vibrations, especially in the
frequency range from 6Hz to 15Hz, can inuence the measured signal.
• Select an installation location which is vibration-free.
• If necessary, change the measuring frequency settings:
– Make sure that the measuring frequencies do not correspond to the fundamental of the
installation site nor one of its harmonic multiples.
– You must also ensure that the measuring frequency diers from half the basic frequency
of the vibration.
Note
Magnetic radiance of the OXYMAT 7
Due to its functional principles, the OXYMAT 7 emits magnetic stray elds. Therefore, do not
operate magnetically sensitive devices in the immediate vicinity of the analyzer module.
Depending on the sensitivity, a distance of up to 50 cm from the analyzer module is required.
5.4.3 Operation with at least one OXYMAT 7
Note
Additional clamping screws
Additionally required clamping screws are not included in the scope of delivery. Order the
clamping screws separately.
Combined operation
5.4Application planning
Wall-mounted device
48 Operating Instructions, 07/2023, A5E31930403-AB