8.1.3 Installing/removing the sample gas restrictor
Requirements
Before you install/remove the sample gas restrictor, ensure that
• the sample gas line is disconnected from the device,
• you have read the section "Combination operation (Page41)",
• a suitable slot screwdriver is at hand.
The dierence between a sample gas restrictor and a clamping screw is:
– Sample gas restrictor: slotted screw; outlet diameter 0.5 mm
– Clamping screw: Cross-tip screw; outlet diameter 2 mm. For additional information, refer
to the section "Structure of the gas connections (Page43)".
Procedure
The installation, removal and retrotting of clamping screws or sample gas restrictors is limited
to only a few module types. You will nd more detailed information on the mounting position
and the mounting conditions in the section "Structure of the gas connections (Page43)".
1. Install the sample gas restrictor or clamping screw:
Screw the sample gas restrictor or the clamping screw into the outlet. The tightening torque
is 1Nm.
2. Connect the sample gas line again.
3. Remove the sample gas restrictor or clamping screw:
Proceed in reverse order.
Connecting the device
8.1Gas connections
Wall-mounted device
94 Operating Instructions, 07/2023, A5E31930403-AB