Retrotting example "Suction operation, serial" operating mode for OXYMAT7/ULTRAMAT7
When you are using the ULTRAMAT7 and the OXYMAT7 analyzer module in serial suction
operation, you must convert the sample gas restrictor of the OXYMAT7 analyzer module:
Sample gas inlet → Sample gas outlet.
a Sample gas inlet
b Reference gas inlet
c Sample gas outlet/reference gas outlet
① Remove sample gas restrictor from the sample gas inlet of the additional OXYMAT7 analyzer
module
② Install the clamping screw into the sample gas inlet of the additional OXYMAT 7 analyzer module
③ Install the sample gas restrictor into the sample gas outlet of the additional OXYMAT7 analyzer
module
④ Install additional OXYMAT7 analyzer module
⑤ Target conguration: ULTRAMAT7 and OXYMAT7 in serial suction operation
Figure5-6 Retrotting of the sample gas restrictors: Example "Pressure operation, serial"
5.4.8 Enclosure purging
Enclosure purging
Gases can escape from the gas path on a very small scale and accumulate over a longer period
of time to dangerous levels. Therefore, the enclosure must be purged with dry, dust-free air or
inert gas when dealing with ammable, toxic or highly corrosive sample gases.
Purge the wall-mounted enclosure with a ow rate of approx. 1 l/min. Purging of the
enclosure can only be dispensed with if non-toxic gases or gas mixtures below the lower
explosive limit (LEL) are introduced into the device. Collect the gas displaced by the purging
using a suitable mechanism. Dispose of the gas in an environmentally friendly way, for
example, via an exhaust pipe.
Combined operation
5.4Application planning
Wall-mounted device
52 Operating Instructions, 07/2023, A5E31930403-AB