V Series Operating Instructions 8 - 13
Alarms Alarm Responses
8.3.4 Configuring Alarm Levels
1. Select the Alarms tab from the Navigation Area.
2. Select the Responses button.
3. Select the applicable alarm group tab.
4. Select the applicable parameter alarm level radio button.
NOTE: The alarm levels available are based on the parameter. Some
parameters will not have all three levels available, as shown in FIGURE
8-3.
8.3.5 Understanding Alarm Response Settings
Based on the selected alarm responses, several monitor responses may occur when a parameter
alarm occurs. Configure each parameter to use any of the following responses:
• Print on Alarm directs the monitor to send the alarm data to the configured printer.
• Save to Event directs the monitor to mark the event and save it to the events database.
NOTE: Apnea alarms are saved to the events database only when
acknowledged. If multiple, short apnea events occur, but
acknowledged only once, only a single event will be saved to the events
database. To view a complete list of apnea events that occurred but
were not acknowledged, refer to the events log on the central station
the monitor is connected to.
•Nurse Call directs the monitor to send an electronic signal to the Nurse Call system.
NOTE: Nurse Call is not an available alarm response unless it was enabled in
System Setup.
8.3.6 Configuring Alarm Response Settings
1. Select the Alarms tab from the Navigation Area.
2. Select the Responses button.
3. Select the applicable alarm group tab.
4. Select the applicable alarm level.
NOTE: Before setting the alarm responses, refer to "Factory Default Alarm
Settings" on page 8-14 for additional information.
5. Optional - select the applicable parameter alarm check boxes under the headings: Print on
Alarm, Save to Event, and Nurse Call.
6. Select the Cancel or Accept button.
•The Cancel button closes the dialog without saving the settings.
•The Accept button saves the settings and closes the dialog.

Table of Contents

Other manuals for Mindray V21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V21 and is the answer not in the manual?

Mindray V21 Specifications

General IconGeneral
BrandMindray
ModelV21
CategoryMedical Equipment
LanguageEnglish

Related product manuals