Alarm Responses Alarms
8 - 14 V Series Operating Instructions
8.3.7 Factory Default Alarm Settings
8.3.7.1 ECG/Resp Tab
TABLE 8-10 lists the available alarm levels for each parameter in the ECG/Resp tab.
8.3.7.2 Pressure Tab
TABLE 8-11 lists the available alarm levels for each parameter in the Pressure tab.
TABLE 8-10
ALARM LEVELS*
FACTORY
DEFAULT ALARM RESPONSES
ALARM LEVELS
BUTTON HIGH MEDIUM LOW
PRINT ON
ALARM
SAVE TO
EVENT
NURSE
CALL
HR X X*** —Off On Off
Extreme Bradycardia X —— Off On Off
Extreme Tachycardia X ——Off On Off
Pacer Failure to Sense** — X X Off On Off
Pacer Failure to Capture** — X X Off On Off
ST** X X XOff On Off
Resp X X X Off On Off
Apnea X X—Off On Off
* X indicates available alarm level, — indicates alarm level not available
** Only available when patient size is Adult or Pediatric
*** The factory default configuration settings are in bold.
TABLE 8-11
ALARM LEVELS*
FACTORY
DEFAULT ALARM RESPONSES
ALARM LEVELS
BUTTON HIGH MEDIUM LOW
PRINT ON
ALARM
SAVE TO
EVENT
NURSE
CALL
NIBP X X** —OffOnOff
IBP 1- 8 X X** —OffOnOff
* X indicates available alarm level, — indicates alarm level not available
** The factory default configuration settings are in bold.

Table of Contents

Other manuals for Mindray V21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V21 and is the answer not in the manual?

Mindray V21 Specifications

General IconGeneral
BrandMindray
ModelV21
CategoryMedical Equipment
LanguageEnglish

Related product manuals