General NIBP Information Non-Invasive Blood Pressure (NIBP)
14 - 10 V Series Operating Instructions
14.3.9 NIBP Stat Measurements
The NIBP Stat mode allows continuous measurements for a maximum of five (5) minutes or 10
measurements. The number of measurements taken is dependent on the cuff pressure. In the Stat
mode, measurements are taken every two (2) seconds.
Start an NIBP Stat measurement by doing one of the following:
Select the NIBP Stat button in the Quick Functions tab.
Select the NIBP Stat button in the NIBP dialog.
NOTE: If the Stat mode is active, “Intrv: STAT” displays in the NIBP tile.
14.3.10 Stopping a Stat NIBP Measurement
To stop a Stat NIBP measurement cycle before the designated measurement period expires:
Select the Stop NIBP button in the NIBP dialog.
Select the Stop NIBP button on the Quick Functions tab.
Select the Start/Stop NIBP button on the VPS.
NOTE: When the Stat cycle ends, the system returns to its previous
measurement state.
14.4 General NIBP Information
14.4.1 Cuff Inflation Time
If the NIBP does not pressurize the cuff to the target pressure within 60 seconds, the cuff vents to
atmosphere, and the NIBP Message Area alternately displays a retry message.
14.4.2 NIBP Pressure Limit Fail Safe
If the cuff is over-pressurized, the cuff automatically vents to the atmosphere, and the NIBP message
window alternately displays Cuff Over Pressure and Unable to Measure. The maximum inflation
pressures are:
Adult: 330 mmHg
Pediatric: 220 mmHg
Neonate: 165 mmHg
14.5 NIBP Tile Layouts
The NIBP parameter is shown in a digital tile and a list tile format.
14.5.1 NIBP Digital Tile Layout
NIBP digital tile layout has two displaying modes: Single Width and Double Width. The Single Width
mode is the default displaying mode. To see NIBP digital tile layout displayed in Double Width mode,
refer to "Displaying NIBP Digital Tile in Large Font" on page 3-16 for additional information.
The NIBP digital tile displays (Single Width as shown in FIGURE 14-4, and Double Width as shown in
FIGURE 14-5):
1. Tile name
2. Message area
3. Systolic value
4. Diastolic value

Table of Contents

Other manuals for Mindray V21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V21 and is the answer not in the manual?

Mindray V21 Specifications

General IconGeneral
BrandMindray
ModelV21
CategoryMedical Equipment
LanguageEnglish

Related product manuals