Addendum to V Series Operating Instructions 1 - 17
Continuous Cardiac Output (CCO) & Continuous Hemodynamics (Optional) CCO Troubleshooting
1.5 CCO Troubleshooting
The CCO tile message area displays messages indicating the current state of the CCO parameter. If
there is more than one message to display, the messages scroll.
SVI STAT High Alarm Violation Indicates that the current value is either equal to or higher than the set alarm
limits.
SVI STAT Low Alarm Violation Indicates that the current value is either equal to or lower than the set alarm
limit
s
.
MESSAGE DETAILS
MESSAGE/PROBLEM REASON SOLUTION
Dashes “- -” display in place
of numerics.
M
easurement is invalid. Check catheter placement.
VDI mode may be incorrect. Change VDI mode to Vigilance.
Check Vigilance Indicates there is an error on the
E
d
wards device that is not a
physiological alarm.
Look at the Edwards device for more
detailed inf
ormation.